Welch Allyn Medical Diagnostic Equipment 204el Heart Rate Monitor User Manual


 
Printer Functions
Propaq Encore Reference Guide 35 Welch Allyn
Printer Functions
Press SETUP, MORE, PRINTER to display the printer menu and setup window.
NEXT
Selects the next setting in the status window.
CHANGE
Changes the currently selected display setting.
PR TREND Prints all trends turned on in the Printer Trend Select Window.
MORE
Pressing the MORE button displays another menu and status window.
PREV MENU
Returns you to the previous menu.
CONTINUOUS Sets the print speed for real time (continuous) measurements to 6.25, 12.5,
or 25 mm/sec. This sets the print speed for a printout obtained by
pressing the START/STOP button on the printer.
AUTO PRINT Automatically prints 8 seconds of patient information every 15
minutes, 30 minutes, 1 hour, 2 hours, or 4 hours. This is the latest
patient information (real time). The print speed is automatically set
to 25 mm/sec.
ALARM PRINT Automatically prints upon an alarm. The Propaq Encore prints 20
seconds of patient information. The first 12 seconds contain
information prior to the alarm. The print speed is automatically set
to 25 mm/sec.
NIBP TICKET Automatically prints an NIBP Ticket when the measurement is
taken.
APNEA TICKET When turned on, an Apnea Ticket is printed at the conclusion of an
apnea alarm and at the one-minute clock interval if the apnea alarm
does not cease.
OXYCRG ON ALARM When turned on, an oxycardiorespirogram will print if an HR/PR,
SpO
2
, or RR/BR alarm occurs. For more information on OxyCRG,
see page 112.
mmH
g
NEXT MORE PREV MENUPR TRENDCHANGE
MCO
2 BR
S
p
O
2
PA
ART
HRT1NIBP
18:45:28
III
1
mV
cm
MON
PRINTER SETUP PAGE
CONTINUOUS : 25.0 mm/s
AUTO PRINT : 15 minutes
ALARM PRINT : OFF
NIBP TICKET
: OFF
APNEA TICKET : OFF
OXYCRG ON ALARM: OFF
102.3
°
F
S
122
35/ 18
( 85)
35
12
92
122
58
80