NIBP
Welch Allyn 60 Propaq Encore Reference Guide
3. Screw the hose connector onto the NIBP connector on the monitor's left side.
4. If motion artifact such as shivering, coughing, or vehicle motion interferes with NIBP
readings, do the following:
Position the patient’s limb away from the body so the applied cuff is not in contact
with the patient’s body or any other object such as a bed rail. Try to keep the cuff at
the same level as the heart.
Connect the ECG leads to the patient and perform ECG monitoring during NIBP.
Set Up the NIBP Channel
Press the NIBP button to display the status window and menu.
A cuff that is not properly connected to the patient may result in a false
reading if the patient and cuff are moved by motion artifact or clinical
personnel during the NIBP measurement. Always verify the cuff is properly
placed on the patient.
START/STOP Starts and stops NIBP measurements. Any time the Propaq Encore is taking a
noninvasive pressure measurement, the START button changes to STOP so you
can stop the measurement in progress. This button initiates the same action as
the START/STOP button at the left side of the screen. Pressing STOP will
automatically vent the cuff.
STOP SMARTCUFINTERVAL TURBOCUFAUTO/MAN
ART
HRT
1NIBP
18:46:00
18:45:28
III
1
mV
cm
MON
NIBP ADULT
MODE : TURBOCUF
TIME : 4:21
SMARTCUF : ON
102.3
°
F
S 122
( 87)
85
109
72
122
58
80
85 50 100 150 200
85
122
58
NIBP
99
S
p
O2
mmH
g
mmHg
122
58
85
This symbol indicates the NIBP reading was taken
in the presence of high motion artifact while
monitoring ECG. Artifact can affect accuracy.
To help reduce artifact, see page 60.
NIBP
122/
58
( 85)
18:45
Manometer Bar.
Systolic, diastolic,
and mean values
are indicated above
the manometer as
small triangles.
Time of last measurement
Last measurement
Measure-
ment mode
Alarm Limits Bell
NIBP values and units
of measure rotate in
and out of this window
for up to one hour
(International English
only)
NIBP values are
displayed in one of
these windows if
vacant. After 16
minutes, the display
changes to that
shown below for 44
minutes.
Time
remaining in
Turbocuf
Mode or Auto
Mode interval