Welch Allyn Medical Diagnostic Equipment 204el Heart Rate Monitor User Manual


 
Printer Functions
Welch Allyn 36 Propaq Encore Reference Guide
PRINTER FAULT Messages
These PRINTER FAULT messages can appear in an equipment alert window.
The front panel of the printer lets you control the basic printer functions.
If you press
FREEZE prior to pressing SNAPSHOT, the printer prints the 8 seconds of patient
information obtained prior to when you pressed
FREEZE.
LOW BATTERY,
PRINTER DISABLED
This message appears when the Propaq’s battery voltage is
less than 7.6 volts. To continue operation, plug the ac power
adapter into the Propaq.
CHECK DOOR The door on the bottom of the printer is open. Close door to
remove this message.
PAPER OUT T
o add printer paper, see page 128.
OVERHEATING The printer is overheating. Service may be required.
START
STOP
SNAP
SHOT
PRINT
TRENDS
Manually starts and stops a printout of
patient information as it is monitored
(continuous or real time).
Hold down top two keys simultaneously to
generate a paper feed.
Hold down top
and bottom keys
simultaneously to
generate a
test strip.
Prints all trends that are enabled in the
Printer Trend Select Window.
Prints the last 8 seconds of data for
nonrespiration waveforms and 32
seconds of compressed waveform history
for respiration waveforms.