Emerson Process Management NGA2000 HFID Fitness Equipment User Manual


 
Instruction Manual
748297-E
June 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Troubleshooting 4-9
Model NGA2000 HFID
Air Baffle (Ref)
102260
See Detail
Sample Jet
657005
Gasket
102256
Jet Assembly
657012
Gasket
102273
Nut, Jet
657016
Air Baffle
.06
Position with tab over
the air hole
(approximately 1/16” to
the right).
End bent to raise it above the tab a
distance equal to the hei
g
ht of the tab.
5. Holding burner base, lift burner cap
off of assembly, set aside.
6. Remove gasket.
7. Refer to Figure 4-7B on page 4-8.
Holding burner base, lift combustion
chamber off, set aside.
8. Refer to Figure 4-8 below. Lift air baf-
fle out of burner base.
9. Remove the sample jet and gasket
from the bottom of the burner base.
10. Remove the jet nut. Grasp jet as-
sembly and lift out (along with upper
gasket) of burner base. Remove bot-
tom gasket.
Figure 4-8. Burner Jets