Emerson Process Management NGA2000 HFID Fitness Equipment User Manual


 
Instruction Manual
748297-E
June 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Installation
2-9
Model NGA2000 HFID
2-6 INSTALLATION CONSIDERATIONS
CHECKLIST
Verify the following:
The Analyzer’s location should be:
Clean
A well ventilated area
weatherproofed
Non-hazardous
Vibration-free
Have stable ambient temperature
The gas cylinders should be equipped
with a clean, hydrocarbon free two stage
regulator and shut off valve.
All external tubing, regulators, valves,
pumps, fittings, etc. are clean.
The correct fuel type is being used.
The THC content of the supply gases
are compatible with the analysis range.
The calibration background gases are
similar to the sample.
The purge air out, burner exhaust, and
bypass are vented to atmospheric pres-
sure. The pressure should be constant.
The burner exhaust tube must be
slanted down a minimum of 6 degrees
from horizontal.
The bypass line connection must be
slanted down a minimum of 6 degrees
from horizontal for drainage of water
condensation.
If required, thermal insulation around the
bypass fitting to prevent condensation in
the bypass restrictor.
If required, thermal insulation for the
sample inlet connection to minimize the
cold spot.
The heated line is at the correct tem-
perature.
The sample, zero, and span gases are
at the correct temperature.
The heated line to have over tempera-
ture protection.
The sample, bypass, and burner ex-
haust tubing material must handle high
temperature and have thermal insulation
to protect from burns.
The purge air out tubing to be 3/8 inch
and less than 4 feet in length.
All external gas connections have been
leak checked.
The dead volume for external sample
and fuel lines have been minimized.
The stainless steel tubing used for the
fuel and sample lines is clean.