Model NGA2000 HFID
Oven Assembly
659551
4-2 COMPONENT REPLACEMENT
a. Oven
Though the oven can be replaced as a
complete unit, all internal components are
field replaceable.
Oven Removal
Refer to Figure 4-3 below, disconnect the
oven's three gas lines and seven electri-
cal cables, noting location of mating con-
nectors for re-installation.
Note
DO NOT remove the fittings from the
gas lines on the detector.
Remove the two hex nuts securing the
oven to the chassis and the two screws
securing oven to the rear panel. Lift oven
assembly from analyzer.
Figure 4-3. Removal of Oven from Chassis