Service Manual Repair procedures 71
5. Re-insert the rubber gasket (part of the fastener kit) into the groove in the front case.
Begin inserting the gasket at the top of the rear case in the four-turn groove area.
Position the rubber gasket flat in the groove area of the front case.
6. Insert the ECG sensor cable into the J1 connector and lock the latching piece.
Caution Do not to bend or tilt any of the neon bulbs. They must remain vertical.
Smooth curve in flex cable
Correct Incorrect
Sharp bend in flex cable
Four-turn groove area
ECG sensor flex cable at J1
Neon bulbs