38 Repair procedures Welch Allyn Micropaq Monitor
2. Slide the SpO
2
board out from under the two bracket hooks (shown at left in the
illustration below).
3. Remove the SpO
2
board.
Removing the rear case
1. Remove the ECG flex cables from connector J1 on the main board.
2. Remove the SpO
2
flex cable from connector J9 to separate the front and rear cases.
Bracket hook
Bracket hook
Caution Do not separate the main board, the subframe, and the radio card.
They are not serviceable. Separation will damage the single-use antenna
connector, which can be replaced only at the factory.
J1 flex cable
J1
J9 flex cable
J9