24 Functional verification Welch Allyn Micropaq Monitor
Resetting the radio to factory defaults
The menu may be entered from either the Network Status screen or the Network Monitor
screen.
1. Press the scroll down key to access the Main menu, and then repeatedly press
until SYSTEM INFORMATION MENU is highlighted.
2. Press the select key to display the SYSTEM INFORMATION MENU screen.
3. Press scroll down key repeatedly until NETWORK STATUS MENU is highlighted.
4. Press select key to display the NETWORK STATUS MENU screen.
5. While the Network Status screen is displayed, press and hold the scroll down key.
While holding the scroll down key press the scroll up key. The Reset Radio menu
screen appears:
6. To reset the radio card to its default settings use the scroll down key to select
YES and press the select key.
7. To exit the Reset Radio menu without resetting the radio card use the scroll keys to
select NO and press the select key.
If “YES” was selected in step 6, the Micropaq monitor waits for the Radio Card to
complete its reset operation and then the Micropaq powers itself off. Remove and
reinsert the battery to restart the unit.
If “NO” was selected the monitor returns to the Network Status screen.