Welch Allyn Medical Diagnostic Equipment Model 408 Heart Rate Monitor User Manual


 
Service Manual Repair procedures 51
5. Re-insert the rubber gasket (part of the fastener kit) into the groove in the front case.
Begin inserting the gasket at the top of the rear case in the four-turn groove area.
Position the rubber gasket flat in the groove of the front case.
6. Insert the ECG sensor cable into the J1 connector and lock the latching piece.
7. Insert the SpO
2
flex cable into the J9 connector and lock the latching piece.
Caution Do not to bend or tilt any of the neon bulbs. They must remain vertical.
Four-turn groove area
ECG sensor cable
J1
Neon bulbs
SpO
2
flex cable
J9