Invacare RPA600-1E Fitness Equipment User Manual


 
SECTION 8—ACCESSORIES
Electric Portable Patient Lift - RPA6001E 36 Part No 1078988
NOTE:Stablebeingdefinedastheweightfluctuatingtwotenthsofapound.Forexample,a
patientweighingonehundredpounds,thescalewillfluctuatebetween99.8and100.2untilthe
LOCKkeyispressed.Fluctuationoftheweightdisplayedisnormalasnotedabove.Pressthe
LOCKbuttontolocktheweight.
NOTE:TheUNITSbuttoncanbepressedtotogglebetweenunitsofpoundsandkilograms.This
isindicatedbylborkgappearinginthedisplaywindow.
8. Theliftmaynowbeloweredandtheslingremovedfromthepatient.
NOTE:Thepatientʹsweightwillcontinuetobeseeninthedisplaywindow.Thedisplaywillturn
offautomaticallyafteratwominuteperiodofnonuse[nochangesinweightexceedingfivepounds
(twokilograms)].YoucanNOTadjustthetimedelayforautomaticshutoff.Afterthedisplayhas
turnedoff,theweightmayberecalledbypressingtheON/OFFbutton.Theunitcanbeturnedoff
bypressingtheON/OFFbuttonasecondtime.
Replacing the Battery
NOTE:Forthisprocedure,refertoFIGURE 8.4.
NOTE:Thescaleispoweredbyaninevoltalkalinebatterythatshouldprovideapproximately1500
readingsbeforeneedingreplacement.
Whenbatteryreplacementisneeded,LOBATwillappearonthedisplay.Performthe
following:
1. Slidethebatterydooropeninthedirectionofthearrow.
2. Removeexistingbattery.
3. Installthenewbattery.
4. Reinstallthebatterydoor.
FIGURE 8.4 Replacing the Battery
Calibrating the Reliant Scale
NOTE:TheReliantScalewillbeprecalibratedatthefactorywiththeloadcell.Shoulditbe
necessarytorecalibratethescale,followtheinstructionsoutlinedbelow.
1. Thepatientandtheslingmustberemovedfromthescaletoproperlycalibratethe
ReliantScale.Forremovingthepatientinstructions
refertoOperationonpage 14.For
slingdetachmentinstructionsrefertoLiftingthePatientonpage 19.
2. WiththeReliantScaleon,removethefourscrewsonthebackoftheenclosureand
removethefrontcovertoexposethePCboard.
Battery
Door
9V Battery
Assembled
Scale
Arrow