SECTION 3—OPERATION
Part No 1078988 17 Electric Portable Patient Lift - RPA600-1E
Activating a Mechanical Emergency Release
Primary Emergency Release
NOTE:Forthisprocedure,refertoFIGURE 3.3.
NOTE:Thisprocedurewillbringtheboomdownorliftitupifthehandcontrolisnotfunctioning
properly.
Toactivatetheprimaryemergencyrelease,insertapenintotheholelabeledEmergency
UporEmergencyDownonthecontrolbox.
FIGURE 3.3 Primary Emergency Release
Secondary Emergency Release
NOTE:Forthisprocedure,refertoFIGURE 3.4.
NOTE:Allpatientliftactuatorsareequippedwithamechanicalemergencyrelease.The
mechanicalreleasewillenabletheactuatortoretractwithoutpower.
NOTE:Usetheprimaryemergencyreleasefirstbeforeusingthesecondaryemergencyrelease
procedure.Thisprocedureshouldonlybeusediftheprimaryemergencyreleaseprocedureisnot
functioningorisunreachable.
NOTE:TheliftMUSTbeunderaloadforthemechanicalreleasetofunction.
Toactivatethesecondaryrelease,pullupontheEMERGENCYringandpushdownon
theboomatthesametime.
FIGURE 3.4 Secondary Emergency Release
EMERGENCY
Emergency
Release
Hole
Control
Box
Pull UP on
EMERGENCY
Ring.
Push DOWN
on Boom