SECTION 1—GENERAL GUIDELINES
Electric Portable Patient Lift - RPA6001E 10 Part No 1078988
Beforelifting apatientfromastationaryobject(wheelchair,commodeorbe d),slightlyraise
thepatientoffthestationaryobjectandcheckthatallslingattachmentsaresecure.Ifany
attachmentisnotcorrect,low erthepatientandcorrecttheproblem,thenraisethepatientand
checkagain.
Duringtransfer,withthepatientsuspendedinaslingattachedtothelift,DONOTrollcaster
baseov erobjectssuch ascarpet, raisedcarpet bindings,doorframes,orany unev ensurfaces
orobstaclesthat wouldcreateanimbalanceofthepatientliftandcould causethepatientlift
totipov er.UsesteeringhandleonthemastatALLtimestopushorpullthepatientlift.
Inv acarerecommendslockingtherearswivelcastersONLYwhenpositioningorremoving
thesling(standassistortransfer)fromaroundthepatient.
Inv acaredoe sNOTrecommendlockingoftherearcastersofthepatientliftwhenlifting an
individual.Doingsocouldcausethelifttotipandendangerthepatientandassistants.
Inv acareDOESrecommendthattherearcastersbeleftunlockedduringlifting proceduresto
allowthepatientlifttostabilizeitself whenthepatientisinitiallyliftedfromachair, bedor
anystationary object.
Transferring the Patient
Beforetransferring,checkthatthe product’sweightcapacitycanwithstandthepa tientʹs
weight.
WheelchairwheellocksMUSTbeinalocke dpositionbeforeloweringthepa tientintothe
wheelchairfortransport.
Performing Maintenance
Regularmaintenanceofpatientliftsandaccessoriesisnece ssarytoassureproperoperation.
Afterthefirst12monthsofoperation,inspectallpivotpoints andfastenersforwear .Ifthe
metalis worn,thepartsMUSTbereplaced.Performthisinspectionevery sixmonths
thereafter.
DONOTovertightenmountinghardware.Thiswill
damagemountingbrackets.
Castersandaxleboltsrequireinspectionsev erysixmonthstocheck fortightnessandwear .