Invacare RPA600-1E Fitness Equipment User Manual


 
SECTION 4—LIFTING THE PATIENT
Part No 1078988 19 Electric Portable Patient Lift - RPA6001E
SECTION 4—LIFTING THE PATIENT
Preparing the Lift for Use
NOTE:AlthoughInvacarerecommendsthattwoassistantsbeusedforallliftingpreparation,
transferringfromandtransferringtoprocedures,ourequipmentwillpermitproperoperationby
oneassistant.Theuseofoneassistantisbasedontheevaluationofthehealthcareprofessionalfor
eachindividualcase.
Positioning the Lift
NOTE:Forthisprocedure,refertoFIGURE 4.1.
NOTE:RefertoGeneralGuidelinesonpage 9inthismanualbeforeproceedingfurtherandobserve
allwarningsindicated.
NOTE:Beforepositioningthelegsofthepatientliftunderabed,makesurethattheareaisclear
ofanyobstructions.
ƽ WARNING
The legs of the lift must be in the maximum open position and the shifter handle
locked in place for optimum stability and safety. If it is necessary to close the legs of
the lift to maneuver the lift under a bed, close the legs of the lift only as long as it
takes to position the lift over the patient and lift the patient off the surface of the
bed. When the legs of the lift are no longer under the bed, return the legs of the lift
to the maximum open position and lock the shifter handle immediately.
1. Withthelegsofthebaseopenandlocked,usethesteeringhandletopushthepatient
liftintoposition.
2. Lowerthepatientliftforeasyattachmentofthesling.
FIGURE 4.1 Positioning the Lift