SECTION 6—TROUBLESHOOTING
Electric Portable Patient Lift - RPA6001E 26 Part No 1078988
SECTION 6—TROUBLESHOOTING
NOTE:Ifproblemsarenotremediedbythesuggestedmeans,pleasecontactyourdealeror
Invacare.
SYMPTOMS FAULTS SOLUTION
Patient Lift feels loose. Mast/Base joint loose.
Tie - Rods are loose.
Refer to Assembly on page 11.
Noisy or dry sound
from pivots.
Needs lubrication. Refer to Lubricating the Lift
on page 28.
Electric actuator fails to
lift when button is
pressed.
Hand-control or actuator con-
nector loose.
Battery low.
RED emergency stop button
pressed IN.
Battery not connected prop-
erly to control box.
The connecting terminals are
damaged.
Electric actuator in need of ser-
vice or load is too high.
Charge batteries. Refer to Charging the Battery
on page 18.
Rotate RED emergency stop button CLOCK-
WISE until it pops out.
Reconnect the battery to the control box. Refer
to Charging the Battery
on page 18.
Replace the battery pack. Refer to Charging the
Battery on page 18.
Refer to Replacing the Electric Actuator
on
page 29. Contact your Dealer.
Unusual noise from
actuator.
Actuator is worn or damaged
or spindle is bent.
Refer to Replacing the Electric Actuator
on
page 29. Contact your Dealer.
Lift arms will not lower
in uppermost position.
Lift arms require a minimum
weight load to lower from the
uppermost position.
Pull down slightly on the lift arms.
Lift arms will not lower
during a power retrac-
tion.
Shoulder bolt at the junction of
the boom and mast may not be
properly installed.
Refer to Checking and Tightening Mast Pivot Bolt
on page 30.