Invacare RPA600-1E Fitness Equipment User Manual


 
SECTION 7—MAINTENANCE
Part No 1078988 29 Electric Portable Patient Lift - RPA600-1E
Detecting Wear and Damage
Itisimportanttoinspectallstressedparts,suchasslings,spreaderbarandanypivotfor
slingsforsignsofcracking,fraying,deformationordeterioration.Replaceanydefective
partsimmediatelyandensurethattheliftisnotuseduntilrepairsaremade.
Cleaning the Sling and the Lift
Theslingshouldbewashedregularlyinwatertemperatureof180°F(82°C)andabiological
solution.Asoftcloth,dampenedwithwaterandasmallamountofmilddetergent,isallthat
isneededtocleanthepatientlift.Theliftcanbecleanedwithnonabrasiv ecleaners.
Replacing the Electric Actuator
NOTE:Forthisprocedure,refertoFIGURE 7.2.
1. Removethenut,washerandshoulderboltthatsecuretheelectricactuatortothemast
mountingbracket.
2. Resttheboomonyourshoulderandremovethenut,bolt,plasticbushingandpinch
guardfromtheboommountingbracket.
3. Removetheelectricactuatorassembly.
4. Reversestepsforinstallation.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
FIGURE 7.2 Replacing the Electric Actuator
Boom
Pinch Guard
Bushing
Mast