Invacare RPA600-1E Fitness Equipment User Manual


 
SECTION 7—MAINTENANCE
Electric Portable Patient Lift - RPA600-1E 30 Part No 1078988
Checking and Tightening Mast Pivot Bolt
NOTE:Forthisprocedure,refertoFIGURE 7.3.
1. Liftupthebackoftherubberbootandslideitoffthemastalongtheboom.
2. Checkthattheboltisthroughthebracketandthelocknutistightandsecure.
3. Ifneeded,dooneormoreofthefollowing:
•tightenlocknutandbackoffthelocknut1/8ofaturn.
•replacethelocknut.
4. Repositiontherubberboot.
FIGURE 7.3 Checking and Tightening Mast Pivot Bolt
Replacing the Swivel Bar
ƽ WARNING
After the first year of use, the hooks of the swivel bar and mounting brackets of the
boom should be inspected every six months to determine the extent of wear. If
these parts become worn, replacement must be made.
NOTE:Forthisprocedure,refertoFIGURE 7.4.
1. Theswivelbarcomesattachedtotheboomwithbolt,nut,washers,andpinchguard.
NOTE:Washersshouldbeplacedagainstbothsidesoftheboommountingbrackettoeliminate
damage.
NOTE:Theswivelbarhasthreehooksonbothendstoattachtheslingthatsupportsthepatient
duringlift.
2. Removeexistinghardwareandreplacetheswivelbar.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
Mast
Spacer
(if equipped)
Locknut
Boom
Bolt
Washer
Rubber Boot
Bracket