Welch Allyn Medical Diagnostic Equipment 206el Heart Rate Monitor User Manual


 
Setting Alarm Limits
Welch Allyn 98 Propaq Encore Reference Guide
Setting Alarm Limits
Adjust Limits with STAT SET
When you want to quickly set all alarm limits, the Propaq Encore can calculate new alarm
limits using the patient’s current values. Press the
STAT SET button in the Alarms Menu (SETUP,
ALARMS, STAT SET). The Propaq Encore activates all alarms and calculates the limits for all
monitored vital signs, except apnea delay. The monitor performs a mathematical function
(addition, subtraction, or multiplication) on the current value of the vital signs to arrive at the
new limits. The formulas for these calculations are shown in the table.
Vital Sign If the Patient’s Vital Sign
Value is
Then Calculated New Lower
Limit is
Then Calculated
a
New Upper
Limit is
a.New alarm limits calculated by STAT SET cannot be outside the allowable alarm limit range. If a new limit
is calculated to be above or below the allowable alarm limit range, it defaults to the maximum or minimum
alarm limit allowed for that vital sign.
Heart Rate HR 99
100-250
HR 251
HR x 0.8
HR - 20
Unchanged
HR x 1.2
HR + 20
250
Pulse Rate PR 99
PR 100
PR x 0.8
PR - 20
PR x 1.2
PR + 20
Invasive
Pressure
Inv Prs 25
26 - 99
Inv Prs 100
Inv. Pressure - 5
Inv. Pressure x 0.8
Inv. Pressure - 20
Inv. Pressure + 5
Inv. Pressure x 1.2
Inv. Pressure + 20
NIBP NIBP 25
26 - 99
NIBP 100
NIBP - 5
NIBP x 0.8
NIBP - 20
NIBP + 5
NIBP x 1.2
NIBP + 20
Respiration
Rate/Breath
Rate
RR/BR 25
26 - 99
RR/BR 100
RR/BR - 5
RR/BR x 0.8
RR/BR - 20
RR/BR + 5
RR/BR x 1.2
RR/BR + 20
Temperature Temp 0˚C Temp - 0.5 Temp + 0.5
SpO
2
SpO
2
0% SpO
2
- 5
(min. limit 50%)
100% (adult and pediatric
mode)
SpO
2
+ 5 (neonate mode)
ETCO
2
ETCO
2
0 mmHg
ETCO
2
2.0 (% or kPa)
ETCO
2
- 5 mmHg
(min. 15 mmHg)
ETCO
2
- 0.7 (% or kPa) (min.
2.0% or 2.0 kPa)
ETCO
2
+ 10 mmHg
ETCO
2
+ 1.4 (% or kPa)
INCO
2
INCO
2
0 mmHg
INCO
2
0 (% or kPa)
Not affected by STAT SET INCO
2
+ 5 mmHg
INCO
2
+ 0.7 (% or kPa)
Apnea Delay Not affected by STAT SET