Welch Allyn Medical Diagnostic Equipment 206el Heart Rate Monitor User Manual


 
Monitor (Environmental)
Welch Allyn 152 Propaq Encore Reference Guide
Monitor (Environmental)
Monitor Environmental Specifications
Characteristic Specification
Operating Temperature 0° to 40° C
Shipping and Storage Temperature -20° to 60° C
Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and Storage Altitude -2,000 to 40,000 ft (-610 to 12,192 m)
Operating Relative Humidity 15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shipping and Storage Relative
Humidity
15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shock 50 g
Vibration, Random 0.02g
2
/Hz from 10 to 500 Hz, ramping down to 0.002g
2
/Hz at 2000 Hz.
Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA
DO-160C, Category C.
Vibration, Sinusoidal 0.10 peak to peak inches 5 to 17 Hz, sloping to 0.01 peak to peak
inches at 55 Hz, then sloping to 0.0001 peak to peak inches at 2000
Hz. Operating 1 hour per axis, 3 hours per test. Designed to meet
RTCA DO-160C, Category N.
Electromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a collateral standard of IEC 601-1, for
electromagnetic compatibility. Designed to meet RTCA DO-160C,
Section 21, Category B. Propaq 200 Series SN EA000225 and higher:
MIL-STD 461D:CE102, CS101, CS114, CS115, CS116, RE102,
RS103 (United States Air Force).
Caution The monitor may not meet performance specifications if it is not
used or stored within these environmental specifications.