Welch Allyn Medical Diagnostic Equipment 206el Heart Rate Monitor User Manual


 
Factory Default Settings
Propaq Encore Reference Guide 159 Welch Allyn
Display Wave Select All waves are on except NIBP
Trend Group NIBP
Alarm Limits All are ON except P2
HR Limits Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute
NIBP Limits - Systolic Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
NIBP Limits - Diastolic Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
NIBP Limits - Mean Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
P1, P2 Limits - Systolic Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
P1, P2 Limits - Diastolic Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
P1, P2 Limits - Mean Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
SpO
2
Limits Adult: 85%, 100%
Ped: 85%, 100%
Neonate: 80%, 95%
RR/BR Adult: 5, 30 BrM
Ped: 10, 45 BrM
Neonate: 10, 60 BrM
TEMP Limits 35.0°, 37.8° C
T Limits 0.0°, 2.8° C
ETCO
2
Limits 25, 60 mmHg (3, 8 for % and kPa)
INCO
2
Limits N/A, 5 mmHg (0.7 for % and kPa)
Apnea Delay Adult/Ped: 20 seconds
Neonate: 15 seconds
PRINTER SETTINGS
Printer Alarm Print OFF
Printer Auto Print OFF
Printer NIBP Ticket OFF
Printer Apnea Ticket ON
Printer Print Speed 25 mm/sec
Printer Auto Trend OFF
Printer Trend Selections NIBP and P1 = ON; all others = OFF
Printer OxyCRG on Alarm OFF
a.Any time you change the Date, Filter, Temp F/C, Decimal, or CO
2
Units setting, the new setting also becomes
the powerup default setting.