Using the Propaq Encore
Propaq Encore Reference Guide 21 Welch Allyn
Powerup Equipment Alert: Program Fault, Settings Lost
If a PROGRAM FAULT: SETTINGS LOST, TIME/DAY RESET equipment alert appears when
you turn on the monitor, the monitor cannot recall the programmed custom settings and
current time and date. This can occur if the battery is drained or after new software has been
installed.
If this occurs, the monitor provides a special sequence of display windows to help you regain
use of your monitor as quickly as possible. Do the following:
1. Connect an ac power adapter to recharge the battery (if the battery is drained).
2. Press any button below the equipment alert screen to acknowledge the alert. The
monitor will display the Mode Setup window (shown on page 33).
3. Press these buttons to select one of the Factory patient modes for use:
After you press
YES, the monitor will display the Time/Day window.
4. Press
NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER to store
the new time and date.
5. Turn off the monitor, then turn it on again so the settings will take effect.
The monitor is ready for use. If you want to store some customized patient mode program
settings, refer to page 33.
If you follow these steps and the equipment alert reappears at powerup, the monitor may
need to be serviced and the battery replaced. Contact a qualified service person.
Factory Adult mode
POWERUP*, YES.
Factory Pediatric mode NEXT, POWERUP*, YES.
Factory Neonatal mode NEXT, NEXT, POWERUP*, YES.
These display screens are only displayed in this order if the PROGRAM
FAULT equipment alert occurs.