Welch Allyn Medical Diagnostic Equipment 206el Heart Rate Monitor User Manual


 
Temperature
Propaq Encore Reference Guide 143 Welch Allyn
Temperature
Temperature Specifications
Minimum Time between automatic
measurements
30 seconds (Auto Mode)
2 seconds (Turbo Mode)
Electrosurgery Interference
Suppression
Included in all models: 202EL, 204EL, 206EL
NIBP Performance Per EN 1060-1, EN 1060-3, and ANSI-AAMI SP10-1992
NIBP Safety Per IEC 601-2-30
Characteristic Specification
Range 0° to +50°C; 32° to +122°F
Displays T1, T2, and
T
Probes Compatible with YSI Series 400 and 700 probes. HP side panel only compatible with
YSI 400 and has HP connector.
Units °C and °F selectable
Channel Accuracy Temperature Range Tolerance
0° to +10°C
>10° to +50°C
+32° to +50°F
>50° to +122°F
±0.2°C
±0.1°C
±0.4°F
±0.2°F
Resolution 0.1°C or °F
Electrosurgery
interference
suppression
Included in all models: 202EL, 204EL, 206EL
Characteristic Specification