Equipment Alerts
Welch Allyn 102 Propaq Encore Reference Guide
If the equipment condition also caused a patient alarm, you will need to first suspend the
alarm tone by pressing
SUSPEND, then take the required action.
The Propaq Encore returns the channel to its settings prior to the equipment alert. If a patient
alarm occurred and you turned off any alarm limits, you will need to turn them back on. You
can see which alarm limits are turned on by pressing
SETUP, ALARMS.
Power-up Equipment Alert: Program Fault, Settings Lost
If a PROGRAM FAULT: SETTINGS LOST, TIME/DAY RESET equipment alert appears when
you turn on the monitor, the monitor cannot recall the programmed custom settings and
current time and date. This can occur if the battery is drained or after new software has been
installed.
If this occurs, the monitor provides a special sequence of display windows to help you regain
use of your monitor as quickly as possible. Refer to page 21 and perform the described steps.
If you follow these steps and the equipment alert reappears at powerup, the monitor may
need to be serviced and the battery replaced. Contact a qualified service person.