Welch Allyn 118 Propaq Encore Reference Guide
2. Within 15 seconds, disconnect the Acuity network cable from either the Propaq
Encore side panel or the bedside jack. If the patient will no longer be monitored with
the Propaq Encore, turn off the monitor to erase trend information.
If you do not disconnect the Acuity network cable within about 30 seconds, the
monitor and Acuity attempt to reconnect and prompt you to confirm the patient ID.
Whenever the monitor is connected to Acuity, the
NET OFF function overrides the FREEZE/
UNFREEZE function for the
NET OFF front panel button. Therefore, you cannot freeze the
displayed waveforms when the monitor is connected to Acuity. Also, you cannot initiate a
Freeze Print at Acuity when connected to Acuity. When the monitor is not connected to
Acuity, pressing the FREEZE/UNFREEZE button freezes or unfreezes the displayed
waveforms.
Printing at Acuity
You can print various waveforms from the Propaq Encore to the Acuity system printer. To
print a waveform shown on the display screen, press
SETUP, ACUITY, SNAPSHOT. If you press the
FREEZE button on the front of the Propaq Encore, the button changes from SNAPSHOT to FRZ
PRNT.
Network Alert Message
When the Propaq Encore is connected to the Acuity system, it constantly exchanges
information with Acuity. If the Propaq Encore detects an interruption in this flow of
information, it displays an alert message: NETWORK FAULT, CHECK ACUITY/DATA
COMM CONNECTION.
If the Propaq Encore displays this message, check the Acuity network cable to be sure it is
plugged in to the side panel and to the bedside jack. If the cable is damaged, replace the
cable.
If the cable appears undamaged and the Acuity system is operating normally, ask your
service personnel to check the network and the Propaq Encore Acuity connector.