Welch Allyn Medical Diagnostic Equipment 202el Heart Rate Monitor User Manual


 
NIBP
Propaq Encore Reference Guide 61 Welch Allyn
NIBP Displayed in Waveform Window
AUTO/MAN
This button switches the mode between Automatic or Manual Mode. The
Manual Mode is the default unless you change it by reprogramming your
Propaq. Measurements can be taken at intervals of 1, 2, 3, 5, 10, 15, 30, and 60
minutes. Press START to initiate a measurement.
INTERVAL
Selects the interval at which NIBP measurements are automatically taken. The
interval you select, ranging from one minute to 60 minutes, is shown on the
display next to the word TIME.
TURBOCUF
Automatically starts NIBP measurements and continues to take as many
measurements as possible within five minutes.
SMARTCUF
Enables or disables the Smartcuf motion artifact filter. NIBP measurements can
still be taken when Smartcuf is off. Artifact may interfere with the accuracy of
NIBP measurements with Smartcuf off.
mmH
g
NIBP SETUPINV PRS S
p
O2/CO2ECG/RESP
MCO
2 BR
HRT1NIBP
18:45:28
III
1
mV
cm
MON
N
I
B
P
60
30
0
102.3
°
F
mmH
g
TIME SINCE READING: 0 min
35
135/ 81 (103)
12
80
NIBP
C
O
2
M
(103)
135
81
S
p
O2
99
mmH
g
NIBP SETUPINV PRS S
p
O2/CO2ECG/RESP
MCO
2 BR
HRT1NIBP
19:02:28
III
1
mV
cm
MON
N
I
B
P
60
30
0
102.3
°
F
SYS DIA MEAN
135 81 (103)
mmH
g
TIME SINCE READING: 17 min
35
12
80
NIBP
18:45
C
O
2
M
135/
81
(103)
S
p
O2
99
NIBP TREND
TIME BRSYS/DIA-MEANHR/PR
BPM Br/MHH:MM
10:35
9:46
9:31
9:20
OFF
OFF
OFF
OFF
106
88
93
91
OFF
OFF
OFF
OFF
135/ 81 103
142/ 83 110
145/ 87 112
126/ 85 91
S
p
O2
% mmH
g
PR
T
1NIBP
10:45:28
N
I
B
P
102.3
°
F
mmH
g
TIME SINCE READING: 10 min
106
NIBP
PRINT OXYCRGNXT TRND
135/ 81 (103)
(103)
135
81
The numerics are shown in large
characters for 16 minutes for each
new measurement taken...
By turning on NIBP in the wave
select window, the NIBP numerics
can be displayed in a waveform
window.
... and then they change to the
smaller characters for 44 minutes.
The numerics are removed after 60
minutes.
If NIBP is the only vital sign being
monitored, the numerics are
displayed in a waveform window
above a trend window.