Welch Allyn Medical Diagnostic Equipment 202el Heart Rate Monitor User Manual


 
ECG/RESP
Propaq Encore Reference Guide 43 Welch Allyn
4. Apply the electrodes to the patient.
5. Support the ECG cable so it does not stress the electrode wires, ECG cable connectors,
or electrodes.
Some electrodes may be subject to large offset potentials due to polarization.
This effect is most likely when dissimilar metals are used for different
electrodes, and may be severe enough to prevent obtaining an ECG trace.
Furthermore, recovery time after application of defibrillator pulses may be
compromised when using electrodes of dissimilar metals. Squeeze bulb
electrodes, even if all of the same metal, are particularly vulnerable to this
effect. Stainless steel needle electrodes are prone to having large erratic offset
drifts, and are not recommended.
(G)=LA (-)=RA (+)=LL
G
+
RA
LA
LL
If using MCL1, select LEAD II on the Propaq,
and use all three electrodes.
Standard 3-lead
Configuration
Marriott
Configuration
MCL 1
One of six “v lead”
electrode placement
sites for the fifth (c) lead.
LL
LA
RA
RL
5-lead Configuration
V6
V1
Neonate 3-lead Configuration