Using the Propaq Encore
Welch Allyn 24 Propaq Encore Reference Guide
Propaq Encore Display
The display shows waveforms, vital sign numeric values, Propaq Encore status, and alarm
information in different windows. Different vital sign numeric values (such as heart rate and
blood pressures) have upper and lower range limits. If the Propaq Encore detects a vital sign
value outside of the Propaq's measurable range, the monitor displays – – – (below the range)
or + + + (above the range) instead of the vital sign value.
The screenspace is reallocated when vital signs are added or removed. By changing the size
of the numeric windows below the heart rate, the Propaq Encore provides the best possible
view of all numerics for vitals signs being monitored.
Warning The Propaq Encore will show + + + for HR numerics between 301-
350 beats per minute. Above 350 beats per minute, it may display incorrectly
low heart rates, due to intermittent picking of R-waves.
Due to differences in software versions and standards required by different
countries, the displays shown in this reference guide may be slightly different
than the display on your Propaq Encore.
mmH
g
GAS COMP PREV MENUSOURCERESPONSE
BR
S
p
O
2
PA
ART
HRT1NIBP
18:45:28
III
1
mV
cm
MON
CO2
GAS COMPENSATION: OFF
RESPONSE : NORMAL
CO2 SOURCE : MAINSTREAM
SWEEP SPEED : 6.25 mm/s
BAROMETER : 762.2 mmHg
102.3
°
F
S
122
35/ 18
( 85)
35
12
92
122
58
60
MCO2
Bells indicate alarm
limit status
Status
window
Waveform
window
Temperature values are
displayed here.
Time of day, caution and
status messages.
Systolic, diastolic, and (if
space permits) mean
pressures are displayed.
Oxygen
saturation is a
percentage
value.
Heart Rate in
beats per
minute.
Heart Rate Source: HR
indicates ECG;
PR indicates blood
pressure or SpO
2
.
Noninvasive blood
pressure values can
appear here.
ETCO
2
concentration Breath Rate
from CO
2