Welch Allyn Medical Diagnostic Equipment 202el Heart Rate Monitor User Manual


 
Power
Propaq Encore Reference Guide 155 Welch Allyn
Power
Monitor Power Specifications
Electromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a collateral standard of IEC 601-1, for
electromagnetic compatibility.
Paper
Short-term Storage Environment
(up to 7 days)
-20 to 40°C; 5% to 80% noncondensing
Long-term Storage Environment
(up to 5 years)
25°C (optimal), 65% noncondensing
Characteristic Specification
Mode of Operation Continuous
Battery Pack Type Sealed, gel-type lead acid
Battery Pack Capacity Monitor only: 8 volts, 3 Ampere-Hours; Monitor with Expansion
Modules: 8 volts, 6 Ampere-Hours
Battery Recharger Circuitry Internal, powered by external power adapter
DC Input Power Required 12 to 28 Volts, 25 Watts
Input Fuse Rating 3A/250V, Type 2AG (0.57x 0.177 in)
Operating Times on Battery
Monitor only
Monitor and SpO2 (Baqpaq)
Monitor with Expansion Module with
printer, SpO2 and CO2 Options
2.5 hours
4.5 hours
3 hours
Battery Recharge Time with
instrument on
Range of 8 hours to 12 hours typical, depending upon product
configuration
Battery Recharge Time with
instrument off
Range of 6 hours to 8 hours depending upon product configuration
Recharge time until monitor is
usable, starting with discharged but
non-faulty battery
2 minutes typically (longer time required before NIBP, printer, and
CO
2
are available)
Characteristic Specification