MESSAGES
52
MESSAGE CAUSE CORRECTION
Set clock.
The battery power has run low
and the Date and Time setting on
the unit has disappeared.
Insert new batteries and reset the
Date and Time from the SET menu.
No saved data.
Set clock.
The battery power has run low
and the ECG data stored on the
unit has disappeared or the clock
has been reset.
Insert new batteries and reset the
Date and Time and other settings.
Error Messages
Please insert new
batteries.
The batteries are worn out.
MESSAGE CAUSE
01
ERROR #
Replace both batteries with new
ones.
CORRECTION
Device memory error.
Cannot read data.
An error occurred while trying to
read ECG data stored in unit
memory.
11
Contact
Omron Healthcare
Customer Service
at 1-800-829-6427.
Device memory error.
Cannot erase data.
An error occurred while trying to
erase ECG data stored in unit
memory.
12
Device memory error.
Cannot write data.
An error occurred while trying to
save ECG data to unit memory.
13
SD memory error.
Cannot read data.
An error occurred while trying to
read ECG data stored on the SD
memory card.
21
Format the SD memory card.
Refer to page 22.
* Copy any important data to a PC
before formatting the SD memory
card as all data on the card will be
erased.
Use a new/different SD memory
card. (Always turn off the unit
before removing or inserting SD
memory cards.)
SD memory error.
Cannot erase data.
22
SD memory error.
Cannot write data.
An error occurred while trying to
save ECG data on the SD
memory card.
23
An error occurred while trying to
erase ECG data stored on the SD
memory card.