MESSAGES
clean 1. There is dirt, blood, or 1. Clean the test area
test lint on the test area. according to instructions.
area
2. Your hand or an object 2. Repeat the test. Keep test
covered the test area while area clear.
turning on the meter.
remove The test strip was inserted Remove the strip. Wait for the
strip before the word insert word insert before inserting
appeared on the display. test strip.
5.6 mmol/l (Example) If the word contrl appears Repeat the test with a new test
contrl after a blood test, your strip and apply a larger drop
8:01am 6-01-95 blood sample was too small, of blood.
smeared, or another drop was
added after the test began.
error 1 The sample was applied Repeat the test with a new
retest before the words apply test strip.
sample appeared.
MESSAGE PROBLEM WHAT TO DO