Philips 861304 Heart Rate Monitor User Manual


 
49
Philips Medical Systems
APPENDICES
Troubleshooting while the FRx is not being used
(green Ready light is not on)
Press the blue i-button to check defibrillator status, and follow any instructions the
device gives.
behavior possible cause recommended action
chirps or
i-button flashes
The battery power is low or the pads
need to be replaced.
The pads may be damaged or the
adhesive dried out.
The pads case may be open.
The defibrillator may have been turned
off without a pads set installed.
The Training Pads II set has been left in
the defibrillator.
The Infant/Child Key may have been
left installed.
The defibrillator has been stored
outside the recommended
temperature range.
The defibrillator has detected an error
during a self-test or cannot perform a
self-test, or the Shock button is
damaged.
Press the blue i-button. Replace the
battery or pads if instructed.
Replace the pads with a new set
and do not open the case until pads
are needed in an emergency.
Make sure the pads case is closed.
Make sure the pads are properly
installed. (See Chapter 2 for
directions.)
Remove the Training Pads II set
and replace it with a set of SMART
Pads II.
Remove the Infant/Child Key.
Remove the battery for five
seconds then reinstall it to start
the battery insertion self-test. If it
fails, insert a new battery to repeat
the test. If it fails again, do not use
the defibrillator. If it passes, store
the defibrillator within the
recommended temperature range.
Contact Philips for service if
needed.
no chirping
and/or
i-button does not flash,
or
no response to
pressing i-button
The battery is missing or completely
depleted.
The defibrillator may have been
physically damaged.
Remove the battery for five
seconds then reinstall it to start
the battery insertion self-test. If it
fails, insert a new battery and
repeat the test. If it fails again, do
not use the defibrillator.
Contact Philips for service.