42
Philips Medical Systems
HEARTSTART FRx 861304 OWNER’S MANUAL
APPENDICES
Patient treatment protocol options
parameter settings default default description
“call EMS” voice
reminder timing
• At power on (when the
user turns on the FRx)
• At power on and at the
start of the first patient
care pause
• At the start of the first
patient care pause
•No reminder
At the start of
the first
patient care
pause.
Provides a voice reminder to
make sure emergency medical
services have been called, at
the start of the first patient
care pause.
shock series 1, 2, 3, 4 1 The automatic protocol pause
for CPR is activated each time
a shock is delivered.*
During the protocol pause,
the FRx does not perform
rhythm analysis.
The length of the protocol
pause after a shock series is
completed is determined by
the protocol pause timer
setting.
shock series interval
(minutes)
1.0, 2.0,
∞ (infinity)
1.0 A delivered shock must occur
within 1 minute of the
previous shock to be counted
as part of the current shock
series.
NOTE: This parameter is only
applicable when the shock series
is not configured to the default 1
shock.
* A shock series begins when a shock is delivered after the FRx is turned on. A new shock series begins after a protocol
pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous shock
exceeds the shock series interval setting.