Philips 861304 Heart Rate Monitor User Manual


 
42
Philips Medical Systems
HEARTSTART FRx 861304 OWNER’S MANUAL
APPENDICES
Patient treatment protocol options
parameter settings default default description
“call EMS” voice
reminder timing
At power on (when the
user turns on the FRx)
At power on and at the
start of the first patient
care pause
At the start of the first
patient care pause
•No reminder
At the start of
the first
patient care
pause.
Provides a voice reminder to
make sure emergency medical
services have been called, at
the start of the first patient
care pause.
shock series 1, 2, 3, 4 1 The automatic protocol pause
for CPR is activated each time
a shock is delivered.*
During the protocol pause,
the FRx does not perform
rhythm analysis.
The length of the protocol
pause after a shock series is
completed is determined by
the protocol pause timer
setting.
shock series interval
(minutes)
1.0, 2.0,
(infinity)
1.0 A delivered shock must occur
within 1 minute of the
previous shock to be counted
as part of the current shock
series.
NOTE: This parameter is only
applicable when the shock series
is not configured to the default 1
shock.
* A shock series begins when a shock is delivered after the FRx is turned on. A new shock series begins after a protocol
pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous shock
exceeds the shock series interval setting.