Philips 861304 Heart Rate Monitor User Manual


 
38
Philips Medical Systems
HEARTSTART FRx 861304 OWNER’S MANUAL
APPENDICES
Accessories specifications
HeartStart SMART Pads II 989803139261
non-shockable rhythm — all
other non-shockable
rhythms
c
450 specificity >95% (88%)
a. From Philips Medical Systems Heartstream ECG rhythm databases.
b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy. Automatic External Defibrillators for Public
Access Use: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New
Waveforms, and Enhancing Safety. Circulation 1997;95:1677-1682.
c. Supraventricular tachycardia (SVT) is specifically included in the non-shockable rhythm class, in accordance with AHA recommendations
b
and the AAMI standard DF80.
rhythm class
ECG test
sample
a
size
meets AHA recommendations
b
for adult defibrillation
observed 90% one-sided
performance lower
confidence limit
category specifications
pads for defibrillation,
pacing, monitoring,
cardioversion
Disposable, adhesive pads with a nominal active surface area of 80 cm
2
each,
provided in a disposable plastic case, and an integrated 48 inch (121.9 cm),
typical, cable. Pads in case are designed to fit into carry cases.
SMART Pads II compatibility defibrillator model adult
patient use
infant/child patient use
FRx*
FR2/FR2+
FR/ForeRunner
MRx/XL/XLT/4000
HS1/OnSite/Home
competitive adapters
yes
yes
yes
yes
no; use M5071A
yes
yes
no, use M3870A
no
manual mode only
no; use M5072A
manual mode only
* Pre-connectable to FRx defibrillator only.
pads shelf life Pads package is labeled with a use-by date of at least two years from date of
manufacture.