FConfiguration
Overview
The Philips HeartStart FRx Defibrillator comes with a factory default configuration
designed to meet the needs of most users. This configuration can only be changed
by using HeartStart Configure version 1.0 or higher, Event Review version 3.2 or
higher, or Event Review Pro 3.1 or higher. This software is for use by trained
personnel. Information about HeartStart data management products is available
online www.medical.philips.com/goto/eventreview. See Appendix A for ordering
information.
Device options
The following table includes the features of FRx operation that are not related to
patient treatment
parameter settings default default description
speaker volume 1, 2, 3, 4,
5, 6, 7, 8
8 The volume of the FRx’s speaker is set to 8,
highest.
auto send periodic
self-test (PST) data
On, Off On Enables the periodic self-test data to be
broadcast through the device's infrared
data port.
ECG out data On, Off On Enables the ECG data to be broadcast
through the device's infrared data port.