36
Philips Medical Systems
HEARTSTART FRx 861304 OWNER’S MANUAL
APPENDICES
energy Using HeartStart SMART Pads II for adult defibrillation: 150 J nominal (±15%)
into a 50 ohm load.
Using HeartStart SMART Pads II with Infant/Child Key inserted: 50 J nominal
(±15%) into a 50 ohm load. Sample pediatric energy doses:
age energy dose
newborn 14 J/kg
1 year 5 J/kg
2 - 3 years 4 J/kg
4 - 5 years 3 J/kg
6 - 8 years 2 J/kg
Doses indicated are based on CDC growth charts for the 50th percentile
weights for boys.*
* National Center for Health Statistics in collaboration with the National Center for
Chronic Disease Prevention and Health Promotion. CDC growth charts: weight-for-age
percentiles, revised and corrected November 28, 2000. Atlanta, GA: Centers for Disease
Control and Prevention © 2000.
charge control Controlled by Patient Analysis System for automated operation.
shock-to-shock cycle time < 20 seconds typical, including analysis.
“charge complete”
indicator
Shock button flashes, audio tone sounds; device is able to deliver a shock as
soon as a shock is advised.
patient care
pause-to-shock time
Quick Shock. 8 seconds, typical, from end of patient care pause to shock
delivery.
disarm (AED mode) Once charged, the HeartStart FRx will disarm if:
• patient’s heart rhythm changes to non-shockable rhythm,
• a shock is not delivered within 30 seconds after the FRx is armed,
• the On/Off button is pressed for one second to turn off the FRx,
• the Infant/Child Key is inserted or removed,
the battery has been removed or is completely depleted, or
• the impedance between pads is out of range.
adult shock delivery vector Via SMART Pads II placed in the anterior-anterior (Lead II) position.
infant/child shock delivery
vector
Via SMART Pads II typically placed in the anterior-posterior position.
category nominal specifications