Invacare BAR750 Camping Equipment User Manual


 
SECTION 5—MAINTENANCE
Part No 1143215 27 BAR750
SECTION 5—MAINTENANCE
Maintenance Checklist
Invacarerecommendsthefollowingmaintenanceproceduresbeconductedatleastonce
everythreeyears.Thisformisprovidedasaguidetohelpwithdocumentation.
Mechanical Inspection and Maintenance
Inspectallbedcomponentsfordamageorexcessivewear.
Visuallyexamineallweldsforcracks.
Inspecttheheadandfootspringsectionsforbending,warpingordamage.
Checkdriveshaftanddriveshaftconnectionsforbending,damageorexcessivewear.
Inspectpulltubesandmountinghardwareforbending,damageorexcessivewear.
Inspectallboltsandrivetstoensurethattheyaresecurelytightenedandfunctioning
properly.
Checkcasterstoensuretheylock,ifapplicable,androllproperly.
Ensurecastersarefreeofdebris.
Electrical Inspection and Maintenance
Inspectallelectricalbedcomponentsfordamageorexcessivewear(i.e.,crackedor
brokenhousings,orworncomponents).
Checkpendant,powerandmotorcordsforchafing,cutsorexcessivewear.
Makesureallplugsarefullyattachedandfreeofdamage.
Checkallfunctions(head,foot,andbedendsraiseandlowerproperly).
Pass/Fail
PASS:Bedandallcomponentshavepassedalloftheaboveinspectioncriteriaandthe
bedisreadyforplacementintoaclient’shome.
FAIL:Bed,oranycomponent,hasfailedanyoftheaboveinspectioncriteria.Tagthe
bedorcomponentwithacompletedescriptionofthefailure(s)andhavethebed
serviced.