Graham Field 765000 Camping Equipment User Manual


 
22 765000-INS-LAB-RevB10
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
IEC60601 test
level
Compliance
Electromagnetic
Environment-Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms150 kHz to
80 MHz outside
ISM bands
a
3 V/m 80 MHz to
2.5 GHz
3Vrms
3V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of this device, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
Pd 2.1=
150kHz to 80MHz
Pd 2.1=
150kHz to 80MHz
Pd 3.2=
80 MHz to 2.5G MHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
b
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
c
,
should be less than the
compliance level in each
frequency range
d
.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.