APPENDIX A: EMC INFORMATION
Guidance and Manufacturer’s Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should
make sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment-Guidance
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Complies
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network.
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The
user of this device should make sure it is used in such an environment.
Immunity Test IEC60601 test level Compliance
Electromagnetic
Environment-Guidance
Electrostatic
Discharge(ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/ burst
IEC61000-4-4
±2kV for power supply
line
±1kV for input/out line
±2kV for power supply
line
±1kV for input/out line
Mains power quality should be that
of atypical commercial or hospital
environment
Surge
IEC61000-4-5
±1kV for differential
mode
±2kV for common mode
±1kV for differential
mode
±2kV for common mode
Mains power quality should be that
of atypical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC61000-4-11
<5 % UT(>95 % dip in
UT)for 0,5 cycle
40 % UT(60 % dip in
UT)for 5 cycles
70 % UT(30 % dip in
UT)for 25 cycles
<5 % UT(>95 % dip in
UT)for 5 sec
<5 % UT(>95 % dip in
UT) for 0,5 cycle
40 % UT(60 % dip in
UT) for 5 cycles
70 % UT(30 % dip in
UT) for 25 cycles
<5 % UT(>95 % dip in
UT) for 5 sec
Mains power quality should be that
of atypical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions, it
is recommended that the device be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic of
atypical location in a typical
commercial or hospital environment.
NOTE: U
T