Calibration Procedure . . . continued
Change Cal Spectral Values
The following procedure is only necessary when you want to calibrate to a
different standard. Reflectance values from 390nm to 710nm can be changed.
To change cal values:
1) At the Calibrate? menu, press [REF]
(yes) key to select calibration.
The copyright and software version
are momentarily displayed.
"READ CAL WHITE" is displayed.
2) Press the [REF] (change) key.
"NEW CAL VALUES?" is displayed.
3) Enter New Cal Values.
* Press the [REF] (yes) key to
indicate Yes, you do want to
Set New Cal Values.
☛
Pressing [FUNCTION] (no) key will advance to
changing aperture procedure (see Sec. 8.6).
* Press the [next] key to advance
to next value.
* Press the [--] key to decrease
value.
* Press the [+] key to increase
value.
4) Press [REF] (next) key once more
after 710nm is displayed to advance
to white cal procedure. A cal procedure must be performed.
CALIBRATE?
FUNCTION SETUP REF
no
yes
X-Rite
COPYRIGHT
1993
Ver XXXX
READ CAL WHITE
FUNCTION SETUP REF
changeexit
NEW CAL VALUES ?
FUNCTION SETUP REF
yes
no
SET 390nm VALUE
next
54.56
SEE SEC. 8.6.
FUNCTION SETUP REF
CAL
STORE
(-)
(+)
( )
DECREASE
INCREASE
390nm
400nm
710nm
6-8