50 INJECTING YOUR INSULIN
InDuo™ insulin doser function check
If you think that the InDuo™ insulin doser is not
working properly, follow this procedure:
Screw on a new NovoFine
®
needle, as described
on page 41.
Prime the insulin doser as described on page 42.
Put the outer needle cap over the needle.
Dispense 20 units into the outer needle cap.
The insulin should fill the lower part of the outer
needle cap.
If too much or too little insulin is released,
repeat the test.
If the InDuo™ insulin doser again releases too
much or too little insulin, do not use your insulin
doser. Do not try to repair a faulty insulin doser.
Call the InDuo™ Customer Care Line at
1 888 729-2299.