13
© Adam Equipment Company 2012
6.5
ULTRASONIC
HEIGHT
CALIBRATION
F
FF
F5
55
5
UL
ULUL
UL
TR
TRTR
TR
A
AA
A
The Ultrasonic height sensor has been factory calibrated and should not require
frequent calibration. If there is a problem with height measuring proceed as below.
1) Switch on the scale then press the [kg/lb] key during the self-checking test of
the display.
2) The display will show the first function
F1
F1F1
F1
EL.
EL.EL.
EL.
3) Press the [kg/lb] key to select
F5
F5F5
F5
U
UU
U
LTRA
LTRALTRA
LTRA
and then [Zero] to enter the sensor
calibration section “AdJUST” XX (XX being a factory set figure)
4) Press the [Zero] key and the display will show “base- H 220”, press [Zero] to
calibrate the “Zero” height reading.
5) The display will show “test” and a 4 digit number. Press the [Zero] key from a
position where the height sensor cannot detect you.
6) When calibration is complete the display will show “AdJUST XX”.
If after calibration the height sensor is not measuring correctly, press [Print] or
[Clear] at point 3 to change XX. A lower number will decrease the height
measurement and a higher number will increase it. Then continue from point 4 on.
*A print out with date and time can be issued following every calibration.
6.6 SETTING
THE
HOLD
FUNCTION
F6
F6F6
F6
HOLD
HOLDHOLD
HOLD
The Hold function can be set so that the height and BMI readings are visible as they
are being calculated (OFF), or remain as dashes until the weight reading is stable
upon where they will be displayed (ON).
1) Switch on the scale then press the [kg/lb] key during the self-checking test of
the display.
2) The display will show the first function
F1
F1F1
F1
EL.
EL.EL.
EL.
3) Press the [kg/lb] key to select
F
FF
F6
66
6
HOLD
HOLDHOLD
HOLD
and then [Zero] to enter and show the
current setting, default is ON.
4) Press the [Print] key to change the setting for Hold, press [Zero] to save, then
[kg/lb] to select the next parameter.