36
Chapter4SelfCalibration
4-2. Setting Monitor Date and Time
Set the monitor date and time.
Note
• When ColorNavigator is started, the date and time are automatically set. For details, refer to the ColorNavigator User’s
Manual (on the CD-ROM).
Procedure
1. Choose “SelfCalibration” from the Adjustment menu, and press
.
2. Choose “Settings” from “SelfCalibration”, and press
.
3. Choose “Clock Adjustment” from “Settings”, and press
.
The Clock Adjustment menu appears.
4. Set date with or , and press .
5. Set time with
or , and press .
Attention
• If the main power supply is disconnected for an extended period of time, the clock may need resetting.
4-3. Setting Calibration Schedule
Set the calibration schedule for SelfCalibration.
Procedure
1. Choose “SelfCalibration” from the Adjustment menu, and press
.
2. Choose “Settings” from “SelfCalibration”, and press
.
3. Choose “Schedule” from “Settings”, and press
.
The schedule setting menu appears.
4. Choose “Starting Time”, and press .
5. Set the “Starting Time” with
or .
Select the timing to execute SelfCalibration upon reaching the time set in the schedule.
Settings Function
Off SelfCalibrationisnotexecuted.
PowerSave
Tobeexecutedunderanyofthefollowingconditions.
• Whenthemonitorisin“PowerSave”modeorpowerisoffatthesettime.
• Themonitorshiftstothepowersavingmodeorthepoweristurnedoffwhen
thetimeperiodsetintheschedulehaselapsed.
Immediately SelfCalibrationisexecutedimmediatelyatthesettime.
Press to exit.
6. Choose “Setting Type, and press .
Select the method for setting the adjustment schedule.
Setting Function Setting Example
Type-1 Executeonceineveryfewmonths.
ExecuteonceamonthonWednesdayat1:00AM
oftherstweek.
Type-2 Executeonceineveryfewweeks.
ExecuteonceeveryfourweeksonWednesdayat
1:00AM.
Type-3
Executewhenthemonitorusagetime
exceedsthesettime.
Executewhenthemonitorusagetimeexceeded
200hours.
Press