A SERVICE OF

logo

3.5LSM-512S
1Specifications
MODEL LSM-512S
Order No. 544-540
Applicable display unit LSM-6200
Laser scanning range mm(inch)
Up to 126 (5.0)
Measuring range mm(inch)
1 to 120 (.04 to 4.72)
Resolution µm(inch)
0.1 to 100 (.000005 to .005) [Selectable]
Repeatability [*1] µm(inch) ±0.85 (±.000033)[*2]
Whole range µm(inch) ±6.0 (±.00024)[*3]
µm
±( 4.0 + 0.5 ΔD )
Linearity
[*1]
Narrow measuring
range
inch
±( .00016 + .00002 ΔD )
[*3][*4]
Positional error [*1][*5] µm(inch) ±8.0 (±.0003)
Measuring region mm(inch)
30 × 120 (1.2 × 4.72)
[Optical axis direction × Scanning direction]
Number of scans for averaging scan 1 to 2048
Laser classification
Class 2 (Max. Output:1.3mW with a scanning laser, semiconductor
laser: wavelength 650nm)
Number of laser scans /sec 3200
Laser scanning rate m/sec(inch/sec)
904 (35590/sec)
Protection level IP64
Distance between the laser
emission unit and reception unit
mm(inch)
Standard 321(12.64)
Max. 700 (27) [*6]
Temperature
0 °C to 40 °C
Humidity 35 %RH to 85 %RH [without condensation]
Operation environment
Altitude 2000 m or less
Temperature
-15 °C to 55 °C
Storage environment
Humidity 35 %RH to 85 %RH [without condensation]
[*1] Environment for accuracy validation: 20°C ± 1°C temperature; 50% ± 10% humidity.
[*2] A value of ±2σ with a 120mm-diameter gage has been measured for two minutes with a measurement interval of
0.32 seconds, where σ is the standard deviation.
[*3] The value of measurements in the center of the measurement region.
[*4] ΔD is the difference in diameter of the workpiece and the master gage.
[*5] Error due to the positional shift of workpiece in optical axis direction or scanning direction.
[*6] The distance between the laser emission unit and reception unit other than the standard, may affect the accuracy .
2Standard accessories
Part No. Item Qt.
ID uni[*7] 1
02AGN770A Signal cable (5m / 16ft) 1
99MBC094A User’s manual 1
[*7] The ID unit stores data unique to the measuring unit. A replacement ID unit is available at extra cost.
3Optional accessories
Order No. Item
02AGD150
Calibration gages set φ 1 mm φ 120 mm
02AGD260 Air-blow unit for LSM-512 (also available for LSM-512)
02AGN780A/B/D Extension signal cables ( 5m/10m/20m)(16ft/32ft/64ft)[*8][*10][*11]
02AGD150A/B Extension connecting cables(1m/3m)(3.2ft/9.6ft)[*9][*10][*11]
[*8] The signal cable can be extended up to 30 m(96ft).
[*9] The connecting cable can be extended up to 5 m(16ft).
[*10] The total length of the signal cable and the intermediate cable can be extended up to 32 m.
[*11] The length of the
signal cable and the connecting cable other than the standard, may affect the accuracy .
No. 99MBC094A
3-10