Servicing
Troubleshooting
To check the line-power fuse
1
Locate the line-input connector on the instruments rear panel.
2
Disconnect the line-power cable if it is connected.
3
Use a small flat-blade screwdriver to open the pull-out fuse drawer.
The recommended fuse is an IEC 127 5
×
20 mm, 6.3A, 250 V, Agilent Technol-
ogies part number 2110-0703. Notice that an extra fuse is provided in a drawer
located on the fuse holder.
WARNING
For continued protection against fire hazard, replace line fuse only
with same type and ratings, (type T 6.3A/250V for 100/240V
operation). The use of other fuses or materials is prohibited.