8 760000-INS-LAB-RevA08
increase pressure under the buttocks). Press the therapy
adjust button at the left end of the row to cycle between
the settings. The selected therapy mode setting's indicator
will illuminate.
Front panel fourth row
Panel lock button: The panel lock protects the panel set-
tings from accidental change. Press to lock; press for three
seconds to unlock. When the panel remains untouched for
three minutes, it locks automatically. When the panel is
locked, the panel lock indicator will illuminate.
Power button (ON/OFF): The power button turns power to
the mattress system ON or OFF. Press to turn ON; press to
turn OFF. When power is on, the power (ON/OFF) indicator
will illuminate. When power is turned OFF, mattress system
will slowly deate. Note: Main power switch on side of pump
must be ON in order for the panel power button to function.
Alarm mute button: The alarm mute button temporarily
suspends the LED indicator and buzzer when either the
low-pressure alarm or power failure alarm is activated.
Should the problem not be resolved within ve minutes,
the alarm will resume.
Low pressure indicator: The low pressure indicator illumi-
nates, and the alarm sounds, when pressure level is lower
than the selected pressure. If the low pressure indicator
persistently illuminates inappropriately:
1. Check to ensure all connections are properly and se
-
curely connected per installation instructions.
2. Check for any leakage (tubes or connecting hoses). If
necessary, contact your Graham-Field equipment pro-
vider to replace any damaged tubes or hoses.
3. If problem persists, contact your Graham-Field equip-
ment provider.
Power failure indicator: The power failure indicator illu-
minates, and the alarm sounds, when there is a power failure.
Depress the alarm mute button to disable both alarm and LED.