Graham Field 760000-INS-LAB-RevA08 Camping Equipment User Manual


 
760000-INS-LAB-RevA08 11
Note: The mattress system will automatically enter Autofirm
mode for the quickest inflation at set-up; initial inflation takes
approximately twenty minutes.
3. The pump will begin delivering air into the mattress.
4. When the selected pressure level is reached, the pump
will enter alternating mode. The mattress can then be
adjusted to the desired pressure.
5. Replace all sheets, blankets, and pillows on the bed.
pressure adjustment
Generally, a lighter patient will need a lower (softer) set-
ting while a heavier patient will need a higher (rmer)
setting, but pressure adjustment must ultimately be based
on the patient's weight distribution. A short, stocky person
may be the same weight as a tall, thin person, but their
settings will not be the same. The mattress should be rm
enough to support the patient, yet soft enough so the
patient will sink down to eliminate pressure points. Initially
inate the mattress to maximum pressure. Then, when
patient lies on the mattress, test the pressure by placing
four ngers below the mattress under the patient's but-
tock area; your ngers should just t. If the t is too loose,
decrease the setting. If your ngers do not t well, increase
the setting two notches rmer. Wait ten minutes and then
check again. You should only have to adjust this setting
one time per patient. Please consult your physician for an
appropriate setting.
quICk dIsConneCt
Pull the power plug from the wall connector to disconnect
the device quickly.