Diamondback Lode Valiant Treadmill Treadmill User Manual


 
1
Hans Rudolph
Fixed Orifice Pneumotachs
Fixed-orifice pneumotachs for use with the RSS 100 Research
Pneumotach system.
Adult Flow Sensor
Range 1.0 to 180 L/min
Accuracy +/- 3% of reading
Resolution 0.1 L/min
Neonatal Flow Sensor
Range 0.24 to 40 L/min
Accuracy +/- 3% of reading
Resolution 0.01 L/min
Order code
171109 Adult Fixed-Orifice pneumotach
171110 Neonatal Fixed-Orifice pneumotach
Hans Rudolph Linear Pneumotachs
Specifically designed, patented and manufactured for Physiological
and Respiratory testing, the Hans Rudolph linear pneumotachs
utilise a unique housing configuration and screen assembly design
to convert the flow of gas in either direction into a proportional linear
signal of differential pressure from the two pressure taps for input into
a differential gas pressure transducer. Models are available with the
appropriate parameters, such as Volume Dead Space, Linear Flow
Range and Large/Small Flange Port tube size connection.
All models are available as Heated and Non-Heated Pneumotachs
and are supplied with an individual calibration record.
Order codes
3830 Adult heated pneumotach (0-400 L/min)
4830 Adult non-heated pneumotach (0-400 L/min)
3813 Adult exercise heated pneumotach (0-800 L/min)
4813 Adult exercise non-heated pneumotach (0-800 L/min)
Many other models available
Hans Rudolph Heater Controls
Proportional solid state heater control accurately regulate temperature.
Order codes
3850AF Heater control for one pneumotach (230v - 50Hz)
3850BF Heater control for two pneumotachs (230v - 50Hz)
Hans Rudolph Pneumotach
Amplifier series 1110
Provides the instrumentation necessary to produce an analogue
output proportional to flow when connected to any Hans Rudolph
linear pneumotach. It will also provide an analogue output of an
independent auxiliary pressure signal for measurement of pressure
anywhere in the circuit.
Features
Flow Ranges from 0-3, 0-5, 0-10, 0-35, 0-100, 0-160,
0-400 and 0-800 litres per minute
Zero and gain adjustment pots for both outputs
LED Flow direction indicator
Order codes
1110A Pneumotach amplifier +/- 70cmH2O range
1110B Pneumotach amplifier +/- 140cmH2O range
Hans Rudolph Research
Pneumotach RSS100
The RSS100 and RSS100/HR research systems measure gas
flow using the differential pressure method. As gas flows through
the pneumotach, the microprocessor based system converts the
measured differential pressure to volumetric flow rate. Flow and
pressure measurements are used to calculate several ventilatory
parameters such as respiratory rate, lung compliance, tidal volume,
minute volume, positive end expiratory pressure, etc.
Features
Custom Windows software for displaying, storing and
reviewing data
Embedded microprocessor calculates flow rates and
provides correction for gas density, viscosity, temperature,
barometric pressure and airway pressure
Flow Ranges from 0-3, 0-5, 0-10, 0-35, 0-100, 0-160, 0-400
and 0-800 litres per minute
The RSS100/HR is for use with linear pneumotachs. Please
order the appropriate Hans Rudolph heated or non-heated linear
pneumotach and heater control unit.
The RSS100 is for use with adult and neonatal fixed-orifice
pneumotachs.
One of each size is supplied with the module.
Additional quantities sold separately.
Order codes
RSS100 Research pneumotach system
RSS100/HR Research pneumotach system
HANS RUDOLPH
21
58