A SERVICE OF

logo

Temperature
Propaq Encore Reference Guide 143 Welch Allyn
Temperature
Temperature Specifications
Minimum Time between automatic
measurements
30 seconds (Auto Mode)
2 seconds (Turbo Mode)
Electrosurgery Interference
Suppression
Included in all models: 202EL, 204EL, 206EL
NIBP Performance Per EN 1060-1, EN 1060-3, and ANSI-AAMI SP10-1992
NIBP Safety Per IEC 601-2-30
Characteristic Specification
Range 0° to +50°C; 32° to +122°F
Displays T1, T2, and
T
Probes Compatible with YSI Series 400 and 700 probes. HP side panel only compatible with
YSI 400 and has HP connector.
Units °C and °F selectable
Channel Accuracy Temperature Range Tolerance
0° to +10°C
>10° to +50°C
+32° to +50°F
>50° to +122°F
±0.2°C
±0.1°C
±0.4°F
±0.2°F
Resolution 0.1°C or °F
Electrosurgery
interference
suppression
Included in all models: 202EL, 204EL, 206EL
Characteristic Specification