Invacare MA80 Camping Equipment User Manual


 
SECTION 2—OPERATION
Part No 1148138 17 MA80 Series
5. Perform one of the following:
Cancel - Press the Fowler button to exit Fowler Transmitter Setup without linking
to a transmitter.
Save - Wait 60 seconds without pressing a key.
Max Inflate Button
1. Press the Max Inflate button ( ) to select the max inflate mode and inflate the
mattress rapidly to maximum firmness (pressurized to 35 ± 6 mmHg). A series of
beeps sound every three minutes as a reminder that the Max Inflate mode is active.
After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the
previous setting.
NOTE: It takes 25-60 seconds for the mattress to inflate fully (inflation time depends on size of
mattress).
NOTE: It is recommended that Max Inflate setting be used during patient ingress/egress, patient
wound care, patient turning or patient cleaning.
2. Press the Max Inflate button again to manually disengage Max Inflate mode.
NOTE: This will deactivate the Max Inflate LED.
Lock/Alarm Silence Button
1. Press the Lock/Alarm Silence button ( ) to select one of the following modes:
Lock Mode - Pressing this button locks out all control unit functions, including the
On button, to prevent any tampering with the settings. Press the button for
approximately 3-5 seconds for the Lock LED to activate.
NOTE: The Lock LED illuminates when in this mode.
Alarm Silence mode - Pressing this button silences the alarm that sounds in the
event of power failure or when the hose is disconnected from the control unit.
NOTE: The Alarm LED illuminates when in this mode.
Power Fail LED
In the event of power outage, an alarm sounds and the Power Fail LED flashes AMBER.
The control unit has internal memory and retains the previous settings during the power
outage.
Low Pressure LED
In the event that the mattress hose disconnects, an alarm sounds and Low Pressure LED
flashes AMBER. Once the low pressure problem is fixed, the control unit resumes
operation in the previously set mode.