Fuse replacement
tool needed: small screwdriver
1. If you suspect a blown fuse,
disconnect the plug from the wall
outlet immediately.
2. Use a small screwdriver to remove
the cover of the fuse holder (fuse
locations are shown in picture at
right).
3. Insert a new fuse of the correct rating (T1A/250V, VDE
approved).
4. Replace the fuse holder cover. Ensure that fuse holder
cover is securely installed.
Air lter replacement
tool needed: small screwdriver
1. Use a small screwdriver to remove
the air lter plate located at rear of
pump (lter location is shown in
picture at right).
2. Remove the lter. The lter is
reusable, if not torn, and can
be washed gently with a mild
detergent and water. Dry the lter before use.
3. Replace the lter and cover. Ensure that lter cover is
securely installed. Replace the air lter regularly if it is
gray, torn, or the environment is dirty.
STORAGE
1. Lay the mattress out at and upside down.
2. Roll from the head end toward the foot end.
3. Stretch the foot-end strap around the rolled mattress to
prevent unrolling.
4. Store mattress and pump in a dust-free environment
with no exposure to direct sunlight.
s NOTICE:Donotfold,creaseorstackthemattress.
fuse
filter